LSRII SORP
The LSRII is a Special Order Research Product (SORP) digital cell analyser equipped with a Violet (405nm, 100mW), Blue (adjustable 488nm, 80mW; maximum output 100mW), Green (532nm, 150mW) and Red (642nm, 40mW) laser, capable of measuring up to 18 fluorescent parameters in addition to FSC and SSC. N.B. The warm up time for the instrument’s lasers = 30min. If you are the first user of the day, but are not based in the JR, please call ahead and we can switch on the instrument for you.
The LSRII has the optional High-Throughput Sampler (HTS) module, offering automated sample acquisition from 96-well plates; the standard mode is capable of processing a 96-well plate in under 30min (< 0.5% carryover).
This instrument was purchased by the NIHR Oxford Biomedical Research Centre (NIHR OBRC).
The fluorochromes listed in the table below are examples of reagents which can be measured in each detector, but does not represent an exhaustive list. Please ask if you are unsure whether a particular reagent can be detected or not.
Violet | Blue | Green | Red |
A. BV785/786 | A. PerCP, PerCP-Cy5.5 | A. PE-Cy7 | A. APC-Cy7, APC-AF750, APC-H7 |
B. BV711 | B. FITC, AF488, BB515 | B. PE-Cy5.5 | B. AF700 |
C. BV650 | C. PE-Cy5 | C. APC | |
D. BV605 | D. PE-CF594 | ||
E. QD570 |
E. PE | ||
F. QD565* | |||
G. BV510 | |||
H. BV421 |
General information
-
Sample acquisition: >15,000 events/sec (2x107 cells/ml running at 60μl/min)
-
The ‘Sample fine adjustment knob’ allows you to adjust the sample flow rate to intermediate levels. When the knob is at its mid-point: LO = 12µl/min, MED = 35μl/min, HI = 60μl/min. Five full revolutions from the mid-point provide 0.5-2X the mid-point rate e.g. sample flow rates of 6-24µl/min are possible when the LO button is selected.
-
Allow at least 5 minutes for the pressure to stabilise after priming the instrument – this is crucial prior to running the CS&T as not doing so could result in erroneously calculated laser delays and suboptimal data.
-
Unless the HTS is attached, under no circumstance may the outer needle be removed when running samples.
LSRII SORP configuration:
Laser |
PMT | Dichroic mirror | Bandpass filter | Range, nm |
Violet |
A B C D E F G H |
750LP 685LP 635LP 600LP 570LP 557LP 505LP - |
785/60 705/40 660/20 610/20 585/42 560/40 515/20 450/50 |
750-815 685-725 650-670 600-620 570-600 557-570 505-525 425-475 |
Blue |
A B SSC |
670LP 505LP - |
685/35 515/20 488/10 |
670-703 505-525 483-493 |
Green |
A B C D E |
740LP 685LP 635LP 600LP - |
780/60 710/50 660/20 610/20 585/42 |
750-810 685-735 650-670 600-620 564-600 |
Red |
A B C |
740LP 685LP - |
780/60 710/50 670/14 |
750-810 685-735 663-677 |
(Image used courtesy of BD Biosciences)