Clinical evaluation of a miniaturized desktop breath hydrogen analyzer.
Duan LP., Braden B., Clement T., Caspary WF., Lembcke B.
UnlabelledA small desktop electrochemical H2 analyzer (EC-60-Hydrogen monitor) was compared with a stationary electrochemical H2 monitor (GMI-exhaled Hydrogen monitor).ResultsThe EC-60-H2 monitor shows a high degree of precision for repetitive (n = 10) measurements of standard hydrogen mixtures (CV 1-8%). The response time for completion of measurement is shorter than that of the GMI-exhaled H2 monitor (37 sec. vs 53 sec.; p < 0.0001), while reset times are almost identical (54 sec. vs 51 sec. n.s). In a clinical setting, breath H2-concentrations measured with the EC-60-H2 monitor and the GMI-exhaled H2 monitor were in excellent agreement with a linear correlation (Y = 1.12X + 1.022, r2 = 0.9617, n = 115). With increasing H2-concentrations the EC-60-H2 monitor required larger sample volumes for maintaining sufficient precision, and sample volumes greater than 200 ml were required with H2-concentrations > 30 ppm.ConclusionFor routine gastrointestinal function testing, the EC-60-H2 monitor is an satisfactory and reliable, easy to use and inexpensive desktop breath hydrogen analyzer, whereas in patients with difficulty in cooperating (children, people with severe pulmonary insufficiency), special care has to be applied to obtain sufficiently large breath samples.